CVE Database · CVE-2018-19014
CVSS v3.1
N/A
EPSS
0.79%
Published
Jan 28, 2019
Modified
Nov 21, 2024
Public PoC / Exploit
All weaponized →No public PoC or exploit code indexed for this CVE.
Links to public security research (Exploit-DB, Nuclei, Trickest, GitHub) for defensive use only.
Description
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
Weaknesses (CWE)
Affected Products (8)
References (4)