Loading vulnerability details…
CVSS v3.1
8.8
EPSS
3.68%
Published
Dec 14, 2020
Modified
May 22, 2025
Public PoC / Exploit (1)
All weaponized →Links to public security research (Exploit-DB, Nuclei, Trickest, GitHub) for defensive use only.
Description
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
CVSS Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:HWeaknesses (CWE)
Affected Products (2)
References (3)